Output Voltage | 0-24 V (A&B) |
Fixed 5V output |
Output Current | 0-500 mA (A&B) |
Fixed Supply > 4 ADC |
Constant Voltage Operation | |
Voltage Regulation | |
Line (120 VAC +- 10%) | <= 0.01% + 3 mV Main |
| <= 5 mV for line Fixed supply |
Load | <= 0.01% + 3 mV Main |
| <= 10 mV (Fixed) |
Recovery Time | 100us |
Ripple & Noise (5 Hz to 1 MHz) | <= 1mVrms (≤5 mV rms for fixed output) |
Temperature Coefficient | <= 300 ppm 0 C |
Constant Current Operation | |
Adjustable Current Range | 5% to 100% (A&B) |
Current Regulation | |
Line (120 VAC +- 10%) | <= 0.2% + 6mA |
Load | <= 0.2% + 3uA |
Current Ripple | <= 3 mA rms |
Metering | |
Display | 2 Analog |
Voltmeter Range | 0 to 25 V |
Voltmeter Accuracy | 2.5% of full scale |
Ampmeter Range | 0 to 600 mA |
Ampmeter Accuracy | 2.5% of full scale |
Overload Protection | Current limiting, Reverse polarity, overvoltage, Short circuit |
Power Requirements | 108 - 132 VAC 60 Hz, 120/220/230/240/VAC, +- 10%, 50/60 Hz version available |
Power Consumption | 165W |
Operating Temperature | 0 0 to 40 0 C <=85% R.H. |
Storage Temperature | 0 0 to 40 0 <=85% R.H. |
Dimensions (HxWxD) | 5.5x11.75x10.975'' (140x298x264 mm) |
Weight | 10.5 lbs. (4.8 kg) |
One Year Warranty |
Accessories |
SUPPLIED: Instruction Manual, Test Leads TL-S (3 pairs), Spare Fuse |
OPTIONAL (for all models): TL-SA Hook up cables. |