DC Volts | |
Ranges | 200mV, 2V, 20V, 200V, 1000V |
Resolution | 100uV, 1mV, 10mV, 100mV, 1V |
Accuracy | +-(0.5% + 1 Digit) |
Overload Protection | 200 mV Range 500V DC + AC Peak, 350V RMS Sine: All Other Ranges: 1200V DC + AC Peak, 800V RMS, Sine |
Input Impedance | 10 M |
AC Volts | |
Ranges | 200V, 750V |
Resolution | 100mV, 1V |
Basic Accuracy | +-(1.5% + 4 Digits) |
Frequency Response | 50 to 500 Hz |
Overload Protection | 500 VDC + AC Peak, 750VRMS, Sine |
Input Impedance | 4.5M |
DC Current | |
Ranges | 200 mA, 20 mA, 200mA, 10A |
Resolution | 0.1uA, 1uA, 100 uA, 10 mA |
Accuracy | +-(1% + 1 Digit): 10A Range: +-(2% + 3 Digits) |
Burden Voltage | 10A and 200mA range: 750mV max.; All Other Ranges: 325mV max. |
Overload Protection | 10A Range: Unfused: All Other Ranges: 250V Fuse |
AC Current | |
Ranges | Does not apply |
Resolution | |
Accuracy | |
Frequency Response | |
Burden Voltage | |
Overload Protection | |
Resistance | |
Ranges | 200 , 2kOhm, 20kOhm, 200kOhm, 2MOhm, 20MOhm, 200MOhm (200MOhm: 2707A only) |
Resolution | 0.1 , 1Ohm, 10Ohm, 100Ohm, 1 kOhm, 10 kOhm, 100 kOhm (100kOhm: 2707A only) |
Basic Accuracy | 200 M Range (2707A only): +-(5% + 10 Dgt.); +-(0.8% + 1 Digit) |
| 20 M Range: +-(3% + 1 Digit) |
Overload Protection | 500 VDC + AC Peak |
Diode Test | Tested at 1.6mA and 3.2V Max. |
Transistor Test (hFE) | Does not apply |
Capacitance | |
Ranges | Does not apply |
Resolution | |
Accuracy | |
Test Frequency, Voltage | |
Frequency | |
Ranges | Does not apply |
Resolution | |
Accuracy | |
Sensitivity | |
Overload Protection | |
Logic Probe (TTL) | |
Logic Threshold | Does not apply |
Minimum Pulse Width | |
Input Impedance | |
Overload Protection | |
Temperature | |
Range | Does not apply |
Accuracy | |
-4 0 to + 900 0 F (-20 0 to 500 0 C) | |
900 0 to + 1400 0 F (500 0 to + 750 0 C) | |
Continuity Tone | Tone Sounds within 100 mS for Resistance Under 100 |
General | |
Display | 3 1/2 digit, high-contrast 0.7'' LCD, 1999 counts |
Sampling | 3 times/sec. (nominal) |
Operating Temp: | 32 0 to 104 0 F (0 0 to + 40 0 C), 0 to 70% RH |
Storage Temp | -4 0 to 140 0 F (-20 0 to 60 0 C), 0 to 80% RH |
Power Requirements | 9V NEDA 1604. |
Battery Life | 200 hours (alkaline) nominal 275 hours (alkaline) Nominal (2703B only) |
Dimensions (HxWxD) | 5.9x3.1x1.3'' (150x79x33 mm) |
One Year Warranty |
Accessories |
SUPPLIED: Battery, Test Leads, Instruction Manual, Thermocouple Probe (2706A only) OPTIONAL: TL-2A Deluxe Test Leads, TL-3 Accessory Tip Kit (for TL-2A), PR-28A High Voltage DMM Probe (40kVDC), TL-130A General Purpose DMM Kit Carrying Case (not included): LC-29B |