Technical Specifications Important Note: All accuracies apply to 5% to 95% full scale of each range |
DC Volt Range | Resolution | Accuracy | Impedance | 2 V 20 V 200 V | 100 µV 1 mV 10 mV | ±(0.05% + 5d) | 10 MΩ | 1200V | 1000V | 100mV | ±(0.1% + 5d) | 1200V | 100mV | ±(0.3% + 5d) |
|
AC Volt Range | Resolution | Frequency | Accuracy | Impedance | 2V | 100 µV | 40Hz to 60Hz | ±(0.75% + 10d) | 10MΩ | 60Hz to 1kHz | ±(1.5% + 10d) | 1kHz to 5kHz | ±(2.5% + 10d) | 5kHz to 10kHz | ±(5.0% + 10d) | 20V | 1 mV | 40Hz to 60Hz | ±(0.75% + 10d) | 60Hz to 1kHz | ±(2.5% + 10d) | 1kHz to 5kHz | ±(5.0% + 10d) | 5kHz to 10kHz | ±(10% + 10d) | 200V | 10 mV | 40Hz to 50Hz | ±(1.0% + 10d) | 50Hz to 60Hz | ±(0.75% + 10d) | 60Hz to 1kHz | ±(2.5% + 10d) | 1kHz to 5kHz | ±(5% + 10d) | 750V | 100 mV | 40Hz to 50Hz | ±(1.0% + 10d) | 50Hz to 60Hz | ±(0.75% + 10d) | 60Hz to 400Hz | ±(2.5% + 10d) | 400Hz to 1kHz | ±(5% + 10d) |
|
DC Current Range | Resolution | Accuracy | Over Load Protection | 2 mA 20 mA 200 mA | 100 nA 1 µA 10 µA | ±(0.75% + 5d) | 2A/250V | 2 A | 100 µA | ±(1.0% + 5d) | 20 A | 1 mA | ±(1.0 % + 10d) | 15A/250V |
|
AC Current Range | Resolution | Frequency | Accuracy | Over Load Protection | 2mA | 100 nA | 40Hz to 50Hz | ±(1.5% + 10d) | 2A/250V | 20mA | 1 µA | 50Hz to 60Hz | ±(0.75% + 10d) | 200mA | 10 µA | 60Hz to 1kHz | ±(2.0% + 10d) | 1kHz to 5kHz | ±(3.0% + 10d) | 2000mA | 100 µA | 40Hz to 50Hz | ±(1.5% + 10d) | 50Hz to 60Hz | ±(1.0% + 10d) | 60Hz to 1kHz | ±(2.0% + 10d) | 1kHz to 5kHz | ±(3.0% + 10d) | 20A | 1mA | 40Hz to 50Hz | ±(1.5% + 10d) | 15A/250V | 50Hz to 60Hz | ±(1.0% + 10d) | 60Hz to 1kHz | ±(2.0% + 10d) | 1kHz to 2kHz | ±(3.0% + 10d) |
|
Resistance Range | Resolution | Accuracy | Over Load Protection | 2 kΩ 20 kΩ 200 kΩ 2 MΩ | 0.1 Ω 1 Ω 10 Ω 100 Ω | ±(0.2% + 6d) | 600V DC or Peak | 20 MΩ | 1 kΩ | ±(1% + 10d) |
|
Frequency Range | Resolution | Accuracy | Over Load Protection | 20 kHz | 1 Hz | ±(1.5% + 5d) | 600V DC or Peak | 200 kHz | 10 Hz | ±(2.0% + 5d) |
Frequencies under 500Hz requires a minimum amplitude of at least 4Vpp, other sensitivities are not specified. * The digit error (less than 5 digits) is added in measuring value. |
Audible Continuity Test Volt | Threshold | Over Load Protection | 3V | Less than 200 Ω | 600V DC or Peak |
|
Diode Test Test Volt | Max. Test Current | Over Load Protection | 2V | Approx. 1.0 mA | 600V DC or Peak |
|
Dimensions and Weight |
Dimensions (H x W x D) | 9.4 x 3.5 x 10.6" (240 x 90 x 270mm) |
Weight Approx. | 4.6lb (2.1 kg.) |
Equipment Ratings |
Plug and Socket | 3 wire ac power plug and 3 wire outlet |
Power & Fuse Ratings Input Voltage | Fuse | Power Max. | 103 ~ 126V AC (50/60Hz) | F 0.5A / 250V | 10 W | 206 ~ 252V AC (50/60Hz) | F 0.2A / 250V |
|
Operating Environment Temperature | 0°C to +40°C (Accuracy Specified at 23°C ± 5°C) | Temperature Coefficient | 0.1 x (Specified Accuracy) per °C | Humidity | up to 85% RH (Relative Humidity) to 40°C without temperature extremes causing condensation within the instrument. |
|
Storage Environment Temperature | -20°C to +70°C | Humidity | below 85% RH |
|
Insulation Category II | Portable equipment of local level. |
Pollution Degree | 2 |
Protection to IEC 529 | Ordinary |
Note: Specifications are subject to change without notice. |